Norme internationale
Clinical laboratory testing and in vitro diagnostic test systems — Susceptibility testing of infectious agents and evaluation of performance of antimicrobial susceptibility test devices — Part 2: Evaluation of performance of antimicrobial susceptibility test devices against reference broth micro-dilution
Publié 23 déc. 2021
This document establishes acceptable performance criteria for antimicrobial susceptibility test (AST) devices that are used to determine minimum inhibitory concentrations (MIC) of bacteria to antimicrobial agents in medical laboratories. This document specifies requirements for AST devices and procedures for assessing performance of such devices. It defines how a performance evaluation of an AST device is to be conducted. This document has been developed to guide manufacturers in the conduct of performance evaluation studies.
Contains information from ISO Open Data (https://www.iso.org/open-data.html), made available under the ODC Attribution License (ODC-By) v1.0.
Retirée · 14 juin 2007